Energizer MMT-511 Betriebsanweisung Seite 156

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 170
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 155
144 Chapter 12
Insulin delivery
Allows up to 48 individual basal rates to be programmed in a
daily pattern.
Different basal rates can be programmed every 30 minutes.
Rates can be set from 0.0 to 35.0 units /hour in 0.1-unit
increments (limited by maximum basal rate setting).
The basal delivery capacity is 0.0-35.0 units/hour.
Block
The block feature restricts access to all programming except
suspend, selftest and the delivery of a bolus with the remote
control.
The remote control must be used to program a bolus. The
pump can be Suspended from either the remote or the pump.
Delivery accuracy
The pump has a delivery accuracy of +/- 5%.
Displacement accuracy
The pump has a displacement accuracy of +/- 2%.
Drive motor
The pump motor has a unique, patented design which delivers
in precise increments.
Dual wave bolus
Delivers a normal bolus followed by a square wave bolus
(limited by maximum bolus).
Seitenansicht 155
1 2 ... 151 152 153 154 155 156 157 158 159 160 161 ... 169 170

Kommentare zu diesen Handbüchern

Keine Kommentare